Development and validation of liquid chromatographic method for the determination of nitrofuran residues in fish samples
pdf

Keywords

chromatographic method

How to Cite

Gotsiridze, . D., Baramidze, K., Chikviladze, T., Otarashvili, T., & Ioramashvili, H. (2021). Development and validation of liquid chromatographic method for the determination of nitrofuran residues in fish samples. Collection of Scientific Works of Tbilisi State Medical University, 54, 42–45. https://doi.org/10.52340/csw.2020.01.417

Abstract

Sensitive, accurate and efficient Liquid Chromatographic Method is developed for the quantitative determination of Nitrofuran residues in fish samples. It was found that the coefficient of variation for all four test substances of the Nitrofuran group is from 0.03 ìg / ml less than 2%, respectively the detection minimum was considered to be 0.03ìg / ml; 376 nm wavelength is specific for Nitrofurans; For the quantification of nitrofuran residues in a fish sample validation of the developed Liquid Chromatographic Method revealed complete compliance of the method (Guidance for Industry Bioanalytical Method Validation U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER) Center for Veterinary Medicine (CVM) May 2001) with the following validation characteristics of the requirements: detection minimum, specificity, accuracy, correctness and linearity.

https://doi.org/10.52340/csw.2020.01.417
pdf

References

საქართველოს მთავრობის დადგენილება ¹499 "ტექნიკური რეგლამენტი-ცოცხალ ცხოველებსა და ცხოველური წარმოშობის სურსათში ზოგიერთი ნივთიერებისა (სუბსტანციის) და მათი ნარჩენების გამოკვლევისათვის ანალიზის მეთოდების განხორციელებისა და შედეგების ინტერპრეტაციის წესი”. 08.11.2016.

Commission Decision (2002): Commission Decision 2002/657/EC of 12 August 2002 implementing Council directive 96/23/EC concerning the performance of analytical methods and the interpretation of results. Official Journal of the European Communities, L221, 836. http://faolex.fao.org/docs/pdf/eur49615.pdf;

Commission Decision (2003): Commission Decision

/181/EC of 13 March 2003 amending Decision 2002/657/EC as regards the setting of minimum required performance limits (MRPRs) for certain residues in food of animal origin. Official Journal of the European Com-munities, L71, 1718. http://eur-lex.europa.eu/pri/en/ oj/dat/2003/l_071/l_07120030315en00170018.pdf;

Commission Directive (2004): Commission Directive 2004/1/EC of 6 January 2004 amending Directive 2002/72/EC as regards the suspension of the use of azodicarbonamide as blowing agent. Official Journal of the European Communities, L7, 4546. http://eur-lex.europa.eu/LexUriServ/LexUriServ.do?uri=OJ:L:2 004:007:0045:0046:EN:PDF;

Kennedy G. (2004): Analytical methods for nitrofurans: Lessons to be learned and new developments. In: Joint FAO/WHO Technical Workshop on Residues of Vet-erinary Drugs without ADI/MRL, Bangkok, Thailand, 8791. http://www.fao.org/docrep/008/y5723e/ y5723e0n.htm#bm23;

USP 40 NF 35, <197>, <851> (2019)

European Pharmacopoea 8Ed (2019)

РУКОВОДСТВО по валидации методик анализа лекарственных средств. Под редакцией:

Н.В. Юргеля, А.Л. Младенцева, А.В. Бурдейна, М.А. Гетьмана, А.А.Малина; (2016)

Guidance for Industry Bioanalytical Method Validation U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER) Center for Veterinary Medicine (CVM) May, 2001;

ICH Q2 A (CPMP/ICH/381/95), Validation of analytical procedure: Methodology, London UK,

Downloads

Download data is not yet available.

Metrics

Metrics Loading ...